Heart rate monitors

;

Heart rate is a physiological response and therefore an indirect indicator of physical activity. Records of heart rate can be used to estimate physical activity intensity or energy expenditure (EE) using the assumption of a linear relationship between heart rate and EE (or oxygen consumption, VO2).

This method is limited by considerable inter- and intra-individual variation in the relationship between heart rate and EE [16 ]. Differences between individuals are mainly related to factors such as age, gender, movement economy and physical fitness, meaning that the heart rate for a given intensity of physical activity is not the same for everyone. Heart rate can also be affected by factors other than physical activity, such as ambient temperature, dehydration, illness or emotional stress, and the relative influence from these factors is greatest during low intensity and sedentary behaviours. As a result, assessment of physical activity using heart rate is best at moderate and vigorous intensity levels and least accurate at lower levels of activity intensity.

Various techniques have been developed to overcome these difficulties, including the FLEX heart rate method, and use of individual calibration of the relationship between heart rate and EE. Heart rate monitors can provide minute-by-minute information relating to the pattern of physical activity in terms of duration, frequency and intensity. Estimates of EE and physical activity intensity depend upon how estimates are derived and the level of calibration. The dimensions of physical activity assessed by heart rate monitoring are described in Table P.3.9.

Table P.3.9 The dimensions which can be assessed by heart rate monitor.

Dimension Possible to assess?
Duration
Intensity
Frequency
Volume
Total physical activity energy expenditure
Type
Timing of bouts of activity (i.e. pattern of activity)
Domain
Contextual information (e.g. location)
Posture
Sedentary behaviour

Electrocardiography (ECG) based heart rate monitors measure the electrical potential on the surface of the skin using electrodes attached to the chest (see Figure P.3.8). This results in a voltage time-series signal, which may be stored as the full ECG waveform (as used in Holter monitoring) for offline processing, or may be processed by the monitor’s inbuilt peak detection algorithm to identify heart beats (the QRS complex in the cardiac cycle). Data are then stored as inter-beat interval time-series or may be further reduced to average heart rate values, say every minute. This final step usually employs a degree of filtering to limit impact of missed and/or ectopic beats. There are therefore three steps involved in deriving a robust estimate of average heart rate; one measurement step and two inference steps.

Optical heart rate monitors work on a different principle of measurement, namely by detecting the fluctuation in blood perfusion of the skin (finger, wrist, earlobe) that occurs during the cardiac cycle by differential absorption of light from a LED to a photodiode (photoplethysmography); the subsequent inferential steps of deriving heart rate in beats per minute are similar in principle to those performed on the ECG signal (peak detection to derive beats, followed by filtering to get a robust average heart rate estimate).

The development of heart rate monitors with internal memory capacity for storing multiple days’ of data in high time resolution has increased the feasibility of this objective measure of physical activity; some do not have visual display of heart rate which further improves feasibility for observational research.

There are a wide range of heart rate monitors which vary in sophistication but in practical terms there are three main types of heart rate monitors used to measure physical activity in population health research:

  1. One type is configured as a chest strap (see Figure P.3.8) which is wirelessly connected to a data logger stored in a watch or directly on the device itself. The chest strap has at least two electrodes which are moistened and securely fastened to the lower chest. The chest strap must be tightly fitted to ensure good contact with the skin and this can be uncomfortable for the participant, who may be asked to wear the device during several days of free-living. The Polar Vantage XL, Polar Accurex II and Accumen Basix Performance are examples of this configuration [31].
  2. Heart rate can also be measured using two or more sticky ECG electrodes, a method commonly used to evaluate patients for cardiac conditions (Holter monitoring). When using this method, it is important to provide extra electrodes and instructions for replacing these to avoid issues with placement and skin irritation. The CamNtech Actiheart is one example [29].
  3. Wrist-based heart rate monitoring using photoplethysmography. This is a more recent technological development and has so far been less used in population research; however wearing devices on the wrist usually has very high participant tolerance. Examples include Scosche Rhythm, Mio Alpha, Fitbit Charge HR, Basis Peak, Microsoft Band, and TomTom Runner Cardio [30].
Figure P.3.8 Wear location of a heart rate monitor chest strap.

Calibration procedures

There are two key issues associated with the use of heart rate to estimate physical activity:

  • The linear relationship between heart rate and EE does not hold at low intensity physical activity.
  • There is considerable inter-individual variation in the relationship between heart rate and EE

Due to inter-individual variability of an individual’s heart rate response to activity, individual calibration is required to increase the accuracy of heart rate data to estimate physical activity [20].

This typically includes collection of heart rate response to a submaximal exercise protocol with a known or measured EE cost profile. The relationship derived between observed heart rate and EE during the exercise test is then applied to predict EE in free-living situations. Determining whether to individually calibrate or not (and how) is a matter of weighing up the likely increased validity of the measurement versus the feasibility of the calibration method.

There are different options for calibration. The most intensive methods are likely to produce more precise estimates but may be perceived as more invasive, take more time for the participant to undertake in the laboratory setting, and are therefore not always feasible for large-scale studies. The choice of calibration method needs to take into account the characteristics of the population being studied and secondly the settings under which a calibration test could be carried out. If necessary existing exercise tests can be adapted or new ones developed for a specific application; the underlying principle is the same in that heart rate data over a reasonable intensity range are needed. Examples of different approaches to individual calibration in descending order of validity but ascending order of feasibility are:

  1. Wide-range graded ergometer test with EE measurement
  2. Wide-range graded ergometer test with estimated EE profile
  3. Moderate-range graded step test with EE measurement
  4. Moderate-range graded step test with estimated EE profile
  5. Low-range walk test with EE measurement
  6. Low-range walk test with estimated EE profile
  7. No dynamic calibration (using only resting heart rate)

The choice of calibration protocol and in particular its intensity range will affect the accuracy of the prediction of EE for each heart rate level [8]. Simplified methods, such as walk or step tests or even no dynamic calibration but simply using sleeping heart rate and other possible proxy indicators of fitness (age, sex) have been suggested, accepting the reduction in precision [4, 7, 8].

Simple methods such as the 8-minute ramped step test with a 2-minute recovery are portable and quick, making them feasible for field studies involving larger numbers of people or where facilities are minimal. The use of only sleeping heart rate is an alternative calibration method and measurements taken overnight explain some variation in the heart rate and EE relationship without the need for resource-demanding calibration procedures but it is not as good as a dynamic test [4]. The relationship between the validity and feasibility of different calibration procedures is demonstrated by Figure P.3.9.

Figure P.3.9 Trade-off between validity and feasibility of calibration procedures.
GXT: graded exercise test; XT: exercise test; SHR: sleeping heart rate.
Adapted from: [4].

Heart rate monitoring is used when estimates of physical activity energy expenditure (PAEE) and time spent in moderate and vigorous intensity physical activity are required. The method also provides estimates of frequency, duration, and timing of these activities. The need for individual calibration for achieving better accuracy may limit use of this method for some larger studies [2].

Studies using group-based calibration equations (rather than individual calibration) suggest that the method is suitable as an objective measure of group-level activity in small-to-large epidemiological studies (say up to around 10,000 participants) but improvements in user acceptability and statistical techniques for inference of true heart rate from potentially noisy measurements of ECG [32] or wrist-based photoplethysmography are needed to make this method applicable to very large studies with >100,000 individuals. It should be noted that accuracy is considerably improved by using individual calibration; this is particularly important for interventions that may have altered fitness, since estimates of PAEE and time spent in different intensities of activity may otherwise be differentially biased with respect to intervention group.

Flex heart rate method

Heart rate monitoring is used when estimates of physical activity energy expenditure (PAEE) and time spent in moderate and vigorous intensity physical activity are required. The method also provides estimates of frequency, duration, and timing of these activities. The need for individual calibration for achieving better accuracy may limit use of this method for some larger studies [2].

Studies using group-based calibration equations (rather than individual calibration) suggest that the method is suitable as an objective measure of group-level activity in small-to-large epidemiological studies (say up to around 10,000 participants) but improvements in user acceptability and statistical techniques for inference of true heart rate from potentially noisy measurements of ECG [32] or wrist-based photoplethysmography are needed to make this method applicable to very large studies with >100,000 individuals. It should be noted that accuracy is considerably improved by using individual calibration; this is particularly important for interventions that may have altered fitness, since estimates of PAEE and time spent in different intensities of activity may otherwise be differentially biased with respect to intervention group.

TEE = EE sleep + EE rest + EE physical activity

EE sleep = BMR

EE rest = REE (for time when HR < FLEX HR)

EE physical activity = EE from regression equation (for time when HR ≥ FLEX HR)

The heart rate Flex point is defined empirically as the average of the lowest heart rate during exercise and the highest during rest [5]. Livingstone et al [17] evaluated the accuracy of the Flex heart rate method to estimate TEE in 36 free-living children using DLW. Individual differences ranged from -16.7% to 18.8%; mean group differences ranged from -9.2% ± 4.5% to 3.5% ± 6.6%. The method has also been validated in various adult populations comparing EE estimates with those measured by whole body calorimetry and DLW. Recent estimates from UK adults were reported to agree with DLW-based estimates of PAEE and TEE with no significant bias irrespective of individual calibration level but with higher accuracy and correlations (range r=0.66-0.76 for TEE) for calibrated estimates [5, 10, 18 , 29].

Limitations of the Flex heart rate method

The key limitations of the Flex heart rate method are that it depends on the assumption that the measured heart rate - EE relationship is reliable in free-living daily activities and that it reflects cardiorespiratory responses to physical activity, but not other physiological responses to external stimuli e.g. caffeine or emotional stress [18].

The individual calibration should be undertaken as close to the heart rate monitoring period as possible since the estimation of energy expenditure is based on the assumption that the measured heart rate - EE relationship is representative for the physiological state of the individual during the monitoring period. This relationship may change as a result of improved cardiorespiratory fitness, changes in body composition or certain conditions/diseases (e.g. anaemia).

Other modelling approaches

The feasibility of using heart rate monitoring without individual calibration to estimate EE using prediction parameters has been investigated and found suitable for ranking individuals in epidemiological studies [21]. Other studies have also shown that it is possible to estimate EE from heart rate using multivariate predictive equations derived from group data in adults [11, 9, 13, 25] and children [19].

Variation in applying different prediction equations has been investigated and individual differences found; group level analysis was found to be satisfactory for some population groups [12]. It is important that prediction equations are derived in studies with adequate numbers (>100) and are representative of the population to be studied. Another approach to analysing heart rate data has been to adjust for fitness and age to increase the validity of using heart rate to predict EE [25].

Absolute heart rate indices

Absolute heart rate values have also been used to distinguish between activity intensities [23], albeit less frequently. The method is based on using a heart rate above a certain threshold, e.g. ≥140 beats per minute, as an approximate measure of moderate-vigorous physical activity (MVPA) in children [22].

Relative heart rate indices

A common method to derive relative heart rate indices utilises a measurement of the individual’s resting heart rate (RHR) or expressing heart rate relative to (usually predicted) maximal heart rate. Four of the most commonly used heart rate indices are:

  • Activity heart rate index (AHR): Absolute heart rate minus resting heart rate [4, 28], e.g. reported as mean across the day.
  • Activity heart rate ratio (AHRR)-25: Binary indicator of intensity defined as heart rate >25% above resting heart rate (ratio HR/RHR>1.25). Daily estimates may be reported as proportion of monitored time spent above this level.
  • Activity heart rate ratio (AHRR)-50: Binary indicator of intensity defined as heart rate 50% above resting heart rate (ratio HR/RHR>1.50). Daily estimates reported as proportion of time above this level.
  • Percentage maximum heart rate: Absolute heart rate divided by age-predicted max heart rate.

The impact of differences in the definition of resting heart rate and how it is measured on the estimation of physical activity has been demonstrated [20].

An overview of the characteristics of heart rate monitoring is outlined in Table P.3.10.

Chest strap monitors must be well fitted in order to provide contact with the skin; this may be uncomfortable when measuring heart rate over several days. Alternatively ECG electrodes can be used to obtain heart rate, however these can also be difficult to wear over longer periods due to problems with placement and skin irritation [1].

Strengths

  • Suitable for most populations
  • Low respondent burden for short wearing times
  • Easy and quick for data collection
  • Non wearing time easily identified
  • Waterproof
  • Relatively cheap
  • Physical activity may be measured at group level
  • Linear relationship between heart rate and EE at moderate-to-vigorous intensity activity

Limitations

  • Relationship between heart rate and EE at low levels of activity is not strong so it will not provide an accurate estimate of activity at this intensity.
  • A degree of individual calibration is required.
  • Heart rate response tends lag after changes in movement and remain elevated after movement stops; this means that heart rate may mask sporadic activity; this is of particular relevance in children [27].
  • Factors other than physical activity affect HR:
    • Anxiety or emotional stress, fear or excitement
    • Ambient temperature which may be an issue in hot climates
    • Caffeine and some recreational drugs
    • High dose of beta blockers is a contraindication to the method
  • Participant burden increases over longer periods
  • Electrodes or chest straps may cause discomfort
  • Not always easy to pick up heart rate, due to interference from household devices (telemetry systems) or a bad connection with the skin or other sources of noise which will have to be considered during data cleaning

Table P.3.10 Characteristics of heart rate monitoring.

Consideration Comment
Number of participants Small to large
Relative cost Moderate
Participant burden High with individual calibration
Researcher burden of data collection High with individual calibration
Researcher burden of data analysis High
Risk of reactivity bias Yes
Risk of recall bias No
Risk of social desirability bias No
Risk of observer bias No
Participant literacy required No
Cognitively demanding No

Considerations relating to the use of heart rate monitors for assessing physical activity are summarised by population in Table P.3.11.

Table P.3.11 Physical activity assessment by heart rate monitor in different populations.

Population Comment
Pregnancy
Infancy and lactation Not suitable.
Toddlers and young children May have difficulty wearing electrodes and chest strap both in terms of having more sensitive skin but also the size of straps and belts of the monitors themselves. There are also small pieces which could be swallowed. Child curiosity could lead to fiddling which interferes with heart rate signal.
Adolescents
Adults
Older Adults Safety may be a concern when conducting exercise testing for individual calibration. Self-paced protocols may be viable alternatives. Dexterity may be an issue when changing electrodes/placing device.
Ethnic groups
Other In obese individuals it may be more difficult to get a good heart rate signal due to adiposity acting as a signal dampener.
  • Participant screening may be necessary to ensure health prior to exercise test
  • Instructions on placement and wearing plus contact details should be supplied
  • Participants can be contacted by phone, SMS or email during the planned assessment to encourage compliance
  • Individuals may not wish to wear a chest strap or ECG electrodes; the method may not be suitable for use in very young children
  • Choose an individual calibration protocol suitable for environment/facilities and that participants will be used to, for example lack of experience on a treadmill may cause additional stress leading to inaccurate representation of workload to heart rate; overground walking or running or an exercise bike test may work better in these circumstances
  • People with sensitive skin may experience skin irritation when wearing ECG electrodes
  • A monitor with display of heart rate on a wrist device may influence behaviour (reactivity), especially on day one
  • If reactivity is suspected, data from day one can be compared with subsequent days
  • Heart rate monitors and a range of different sized chest straps (if system comprises chest straps)
  • A range of hypoallergenic ECG electrodes, suitable for long periods of wear (if system comprises ECG electrodes)
  • The level of calibration will determine equipment required; at its most intensive calibration level, it requires a treadmill (or other ergometer) and a respiratory gas analyser
  • At a lower level of calibration a step or a measured distance for a timed walk test may be used
  • Processing of heart rate data requires considerable researcher time and ideally dedicated computer programming

A method specific instrument library is being developed for this section. In the meantime, please refer to the overall instrument library page by clicking here to open in a new page./p>

References

  1. Andre D, Wolf DL. Recent advances in free-living physical activity monitoring: a review. J Diabetes Sci Technol. 2007;1(5):760-7.
  2. Armstrong N, Welsman JR. The physical activity patterns of European youth with reference to methods of assessment. Sports Med. 2006;36(12):1067-86.
  3. Assah FK, Brage S, Ekelund U, Wareham NJ. The association of intensity and overall level of physical activity energy expenditure with a marker of insulin resistance. Diabetologia. 2008;51(8):1399-407.
  4. Brage S, Ekelund U, Brage N, Hennings MA, Froberg K, Franks PW, et al. Hierarchy of individual calibration levels for heart rate and accelerometry to measure physical activity. J Appl Physiol 2007;103(2):682-92.
  5. Ceesay SM, Prentice AM, Day KC, Murgatroyd PR, Goldberg GR, Scott W, et al. The use of heart rate monitoring in the estimation of energy expenditure: a validation study using indirect whole-body calorimetry. Br J Nutr. 1989;61(2):175-86.
  6. Corder K, Brage S, Mattocks C, Ness A, Riddoch C, Wareham NJ, et al. Comparison of two methods to assess PAEE during six activities in children. Med Sci Sports Exerc. 2007;39(12):2180-8.
  7. Corder K, Brage S, Wareham NJ, Ekelund U. Comparison of PAEE from combined and separate heart rate and movement models in children. Med Sci Sports Exerc. 2005;37(10):1761-7.
  8. Corder K, Ekelund U, Steele RM, Wareham NJ, Brage S. Assessment of physical activity in youth. J Appl Physiol 2008;105(3):977-87.
  9. Dugas LR, van der Merwe L, Odendaal H, Noakes TD, Lambert EV. A novel energy expenditure prediction equation for intermittent physical activity. Med Sci Sports Exerc. 2005;37(12):2154-61.
  10. Ekelund U, Yngve A, Westerterp K, Sjostrom M. Energy expenditure assessed by heart rate and doubly labeled water in young athletes. Med Sci Sports Exerc. 2002;34(8):1360-6.
  11. Hiilloskorpi HK, Pasanen ME, Fogelholm MG, Laukkanen RM, Manttari AT. Use of heart rate to predict energy expenditure from low to high activity levels. Int J Sports Med. 2003;24(5):332-6.
  12. Iannotti RJ, Claytor RP, Horn TS, Chen R. Heart rate monitoring as a measure of physical activity in children. Med Sci Sports Exerc. 2004;36(11):1964-71.
  13. Keytel LR, Goedecke JH, Noakes TD, Hiiloskorpi H, Laukkanen R, van der Merwe L, et al. Prediction of energy expenditure from heart rate monitoring during submaximal exercise. J Sports Sci. 2005;23(3):289-97.
  14. Kowalski K, Rhodes R, Naylor PJ, Tuokko H, MacDonald S. Direct and indirect measurement of physical activity in older adults: a systematic review of the literature. Int J Behav Nutr Phys Act. 2012;9:148.
  15. Leonard WR. Measuring human energy expenditure: what have we learned from the flex-heart rate method? Am J Hum Biol. 2003;15(4):479-89.
  16. Li R, Deurenberg P, Hautvast JG. A critical evaluation of heart rate monitoring to assess energy expenditure in individuals. Am J Clin Nutr. 1993;58(5):602-7.
  17. Livingstone MB, Coward WA, Prentice AM, Davies PS, Strain JJ, McKenna PG, et al. Daily energy expenditure in free-living children: comparison of heart-rate monitoring with the doubly labeled water (2H2(18)O) method. Am J Clin Nutr. 1992;56(2):343-52.
  18. Livingstone MB, Prentice AM, Coward WA, Ceesay SM, Strain JJ, McKenna PG, et al. Simultaneous measurement of free-living energy expenditure by the doubly labeled water method and heart-rate monitoring. Am J Clin Nutr. 1990;52(1):59-65.
  19. Livingstone MB, Robson PJ, Totton M. Energy expenditure by heart rate in children: an evaluation of calibration techniques. Med Sci Sports Exerc. 2000;32(8):1513-9.
  20. Logan N, Reilly JJ, Grant S, Paton JY. Resting heart rate definition and its effect on apparent levels of physical activity in young children. Med Sci Sports Exerc. 2000;32(1):162-6.
  21. Rennie KL, Hennings SJ, Mitchell J, Wareham NJ. Estimating energy expenditure by heart-rate monitoring without individual calibration. Med Sci Sports Exerc. 2001;33(6):939-45.
  22. Simons-Morton BG, O'Hara NM, Parcel GS, Huang IW, Baranowski T, Wilson B. Children's frequency of participation in moderate to vigorous physical activities. Res Q Exerc Sport. 1990;61(4):307-14.
  23. Sirard JR, Pate RR. Physical activity assessment in children and adolescents. Sports Med. 2001;31(6):439-54.
  24. Spurr GB, Prentice AM, Murgatroyd PR, Goldberg GR, Reina JC, Christman NT. Energy expenditure from minute-by-minute heart-rate recording: comparison with indirect calorimetry. Am J Clin Nutr. 1988;48(3):552-9.
  25. Strath SJ, Swartz AM, Bassett DR, Jr., O'Brien WL, King GA, Ainsworth BE. Evaluation of heart rate as a method for assessing moderate intensity physical activity. Med Sci Sports Exerc. 2000;32(9 Suppl):S465-70.
  26. Trost SG. Objective measurement of physical activity in youth: current issues, future directions. Exerc Sport Sci Rev. 2001;29(1):32-6.
  27. Trost SG. State of the art reviews: measurement of physical activity in children and adolescents. Am J Lifestyle Med. 2007;1(4):299-314.
  28. Andrews RB. Net heart rate as a substitute for respiratory calorimetry. Am J Clin Nutr. 1971;24(9):1139-47.
  29. Brage S, Westgate K, Franks PW, Stegle O, Wright A, Ekelund U, et al. Estimation of Free-Living Energy Expenditure by Heart Rate and Movement Sensing: A Doubly-Labelled Water Study. PLoS One. 2015;10(9):e0137206.
  30. Stahl SE, An HS, Dinkel DM, Noble JM, Lee JM. How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough? BMJ Open Sport Exerc Med. 2016;2(1):e000106.
  31. Terbizan DJ, Dolezal BA, Albano C. Validity of Seven Commercially Available Heart Rate Monitors. Measurement in Physical Education and Exercise Science. 2002;6(4):243-7.
  32. Stegle O, Fallert SV, MacKay DJ, Brage S. Gaussian process robust regression for noisy heart rate data. IEEE Trans Biomed Eng. 2008;55(9):2143-51.
;